GW INSTEK GRS-6000A Series

Features

* 50M/30MHz with 100MSa/s and 2 Channles 
* Equivalent Time Sampling of 500MSa/s max. 
* Acquire Mode : Peak Detect, Envelop, Persistence 
* Pre-Trigger Function 0 ~ 10 div 
* ROLL Mode to 100s/div 
* Waveform SAVE/RECALL 10 sets (REF0~REF9) 
* Max. Sweep Rate 10ns/div 
* Panel Setting SAVE/RECALL 10 sets (M0~M9) 
* RS-232C Interface

Description:

 

The GRS-6000A Series is a Real-Time/Digital Storage Oscilloscope, which provides both analog and digital storage features in one product. As original signals correctly represent and quickly update, the GRS-6000A Series allows users simultaneously to savie the waveform. With Equivalent Time Sampling at 500MSa/s, the GRS-6000A Series are capable of reconstructing repetitive waveforms up to 50MHz/30MHz in an accurate and detailed fashion. The waveform storage could be done just by pressing the RUN-STOP button, or through the Single Trigger mode which captures single-shot events. 10 Reference Waveforms can be saved and recalled for waveforms comparison. Numerous functions, such as Average, Pre-Trigger, Roll mode, Peak Detect, Envelop, Persistence, and X-Y mode, are available to meet various measurement applications. The RS-232C Interface and the free Remote Control Software enable monitoring, remote controlling, and storing waveforms in PC.

 

 

ANALOG + DIGITAL STORAGE
GRS-6000A Series Real-Time digital storage oscilloscope provides both analog and digital storage features in one product. It is extremely useful for applications requiring real time measurements (analog mode) and waveform storage functions (digital mode) at the same time. With a maximum 500MSa/s equivalent time sample rate, the GRS-6000A Series is able to accurately reconstruct waveforms up to 50MHz (GRS-6052A) and 30MHz (GRS-6032A).
 
DIGITAL STORAGE
WhenRUN STOP button is pressed or Single Trigger mode is used, waveforms can be conveniently captured. Up to 10 reference waveforms can be saved for waveform comparison. With proven analog technology, functions, such as Average, Pre-Trigger, Roll mode, and X-Y mode, offer the flexibility of digital features.
GRS-6000A Series provides a pre-trigger function, which allows you to observe pre-triggered waveforms up to 10 divisions ahead of the trigger point.
  The low-speed transient events from an input signal can be easily seen under the ROLL Mode. The waveform will roll on from right to left to show input signal continuously with a roll time of up to 100s/division.
  With ALT-MAG function, you can expand waveforms by 5, 10 or 20 times for detailed waveform observation. Both the original waveform and expanded waveform can be shown on the screen simultaneously.
 
RS-232C AND REMOTE CONTROL SOFTWARE
 
When remote control software through the RS-232C interface is active, waveforms on the oscilloscope screen can be displayed and updated on a PC screen. The graphic user interface provides an easy platform for remote control. All waveform data can be converted into BMP, PCX, TIFF, PNG, or JPEG files for documentation and analysis.

 

Specification:

 

 
GRS-6052A
GRS-6032A
CRT 
Type and Acceleration
6-inch CRT, 10kV
6-inch CRT, 2kV
Illumination Front panel control
Z-axis Input Sensitivity : at least 5V 
Polarity : positive going input decreases intensity
Max. input voltage:30V(DC+ACpeak)
Input Impedance : approx. 33kΩ Input Impedance : approx. 47kΩ
VERTICAL SYSTEM
Deflection Coefficient and Accuracy 1mV ~ 2mV/div ±5%, 
5mV ~ 20V/div ±3%, 
14 steps in 1-2-5 sequence
Variable Continuity 2.5 : 1 ~ min. 50V/div
Bandwidth (-3dB) 1mV ~ 2mV/div: DC~7MHz
5mV ~ 20V/div: DC~50MHz
1mV ~ 2mV/div: DC~7MHz
5mV ~ 20V/div: DC~30MHz
Vertical Mode CH1, CH2, DUAL (ALT or CHOP)
Chopper Frequency Approx. 250kHz
Sum or Difference CH1+CH2, CH1-CH2
Invert CH2
Input Impedance 1MΩ ±2%//approx. 25pF
Input Coupling AC, DC, GND
Input Voltage Max. 400V(DC+ACpeak)
HORIZONTAL SYSTEM
Sweep Time 0.2μs/div ~ 0.5s/div, 20 steps
Variable Continuity 2.5 : 1 up to 1.25s/div (uncal.)
Accuracy ±3%, ±5% at x5/ x10MAG. ±8% at x 20MAG
Sweep Magnification x5, x10, x20
Max. Sweep Time 20ns/div (10ns/div uncal) 50ns/div (10ns~40ns/div uncal)
ALT-MAG Function Yes
HOLD-OFF Time Variable
TRIGGER 
Trigger Mode AUTO, NORM, TV
Trigger Source VERT, CH1, CH2, LINE, EXT
Trigger Coupling AC, HFR, LFR
Trigger Slope  “+”or“ - ”polarity
ALT Trigger Yes
Indicator Trigger LED Yes
TV Sync. Separator TV-V“-”, TV-H“ - ”
Trigger Sensitivity
 
GRS-6052A
20Hz~5MHz
5MHz~40MHz
40Hz~50MHz
GRS-6032A
20Hz~2MHz
2MHz~20MHz
20Hz~30MHz
CH1, CH2
0.5div
1.5div
2.0div
Vert-Mode
2.0div
3.0div
3.5div
EXT
200mV
800mV
1V
TV sync. pulse more than 1 div or 200mV (EXT)
External Trigger Input 
 
Input impedance :Approx. 1MΩ//25pF (AC coupling)
Max. input voltage : 400V (DC + ACpeak)
X-Y OPERATION
Input X-axis : CH1 ; Y-axis : CH2
Sensitivity 1mV/div ~ 20V/div
Bandwidth X-axis : DC ~ 500kHz (-3dB)
X-Y Phase Shift <3° from DC ~ 50kHz
DIGITAL STORAGE
Acquisition Digitizer 8 bit ADC x 2
Max. Sampling Rate
 
500MSa/s for equivalent time sampling
100MSa/s for normal sampling
Storage Bandwidth Single shot: DC ~ 25MHz
Repetitive: DC ~50MHz
Single shot: DC ~ 25MHz
Repetitive: DC ~30MHz
Dynamic Range ±5div
Memory Length- Acquisition Memory  2k words/CH x 2, 1k words/CH(equivalent)
Memory Length- Save REF Memory 1k words/CH x 10 with back-up memory (REF0~REF9)
Memory Length- Display Memory 1k words/CH x 4 waveform(max.)
Sweep Time Equivalent: 0.2μs/div ~ 0.5μs/div
Normal Sample: 1μs/div ~ 0.1s/div
Roll: 0.2s/div ~ 100s/div
Sweep Magnification x 5, x 10, x 20
Max.Sweep Time 10ns/div
MAG Interpolation DOTS, LINEAR
ALT-MAG Function Yes
Acquire Mode Sample, peak detect(>25ns), Envelop. Persist, Average(2~256)
Operation Mode Auto, Norm, Single, Single-roll, Roll, X-Y, Run/Stop
Smoothing Function Dot joint ON/OFF selectable
Pre-Trigger Pre-trigger 0 ~10div in 0.02div steps
X-Y Operation X-axis: CH1 Y-axis: CH2
Storage Bandwidth DC~50MHz(-3dB)
Display Resolution H: 100points/div; V: 25points/div; 
X-Y: 25 x 25 points/div
SAVE/ RECALL
Panel Setting 10 sets(M0 ~M9)
Waveform 10 sets(REF0~REF9)
READOUT & CURSOR
Cursor Measurement ΔV, ΔT, 1/ ΔT
Readout Intensity Adjustable
OUTPUT SIGNAL
CH1 Signal Output Voltage : approx. 20mV/div (with 50Ω terminated) 
Bandwidth : 50Hz ~ 5MHz
Calibrator Output
 
Voltage : 0.5V±3% 
Frequency : approx. 1kHz, square wave
INTERFACE
  RS-232C
POWER SOURCE 
  AC 100V/120V/230V±10%, 50/60Hz
DIMENSIONS & WEIGHT
 

275(W) x 130(H) x 370(D) mm; Approx. 8.5kg

 

Accessories:

User manual x 1, Power Cord x 1 
GTP-060A-4: 60MHz (10:1/1:1) switchable passive probe (one per channel)

Optional Accessories
GTL-110  Test lead, BNC-BNC heads
GTL-232  RS232 cable, 9 pin female to 9 pin, Null modem for computer
GTC-001  Instrument cart, 450(W) x 430(D)mm (120V input socket)
GTC-002  Instrument cart, 330(W) x 430(D)mm (120V input socket)

Ordering Information:

GRS-6052A  50MHz, Digital storage + Analog oscilloscope
GRS-6032A  30MHz, Digital storage + Analog oscilloscope

 

 

 

 

 

Location