Specifications |
Accuracy specifications |
Function |
Range |
Resolution |
Accuracy ± ([% of Reading] + [Counts]) |
DC Millivolts |
600.0 mV |
0.1 mV |
0.5 % + 2 |
DC Volts |
6.000 V 60.00 V 600.0 V |
0.001 V 0.01 V 0.1 V |
0.5 % + 2 |
Auto-V LoZ true-rms |
600.0 V |
0.1 V |
DC, 45 Hz to 500 Hz: 2.0% + 3 500 Hz to 1 kHz: 4.0% + 3 |
AC millivolts true-rms |
600.0 mV |
0.1 mV |
45 Hz to 500 Hz: 1.0% + 3 500 Hz to 1 kHz: 2.0% + 3 |
AC volts true-rms |
6.000 V 60.00 V 600.0 V |
0.001 V 0.01 V 0.1 V |
45 Hz to 500 Hz: 1.0% + 3 500 Hz to 1 kHz: 2.0% + 3 |
Continuity |
600 Ω |
1 Ω |
Beeper on < 20 W, off > 250 W; detects opens or shorts of 500 µs or longer |
Ohms |
600.0 Ω 6.000 kΩ 60.00 kΩ 600.0 kΩ 6.000 MΩ 40.00 MΩ |
0.1 Ω 0.001 kΩ 0.01 kΩ 0.1 kΩ 0.001 MΩ 0.01 MΩ |
0.9% + 2 0.9% + 1 0.9% + 1 0.9% + 1 0.9% + 1 1.5% + 2 |
Input characteristics |
Function |
Input impedance (nominal) |
Common mode rejection ratio (1 kW unbalanced) |
Normal mode rejection |
Volt AC |
> 5 MΩ < 100 pF |
>60dB at dc, 50 or 60Hz |
- |
Volts DC |
> 10 MΩ < 100 pF |
>100dB at dc, 50 or 60Hz |
> 60 dB at 50 or 60Hz |
Auto-V LoZ |
~3 kΩ < 500 pF |
> 60dB at dc, 50 or 60Hz |
- |
Ohms |
Open circuit test voltage |
Full scale voltage |
Short circuit current |
To 6.0 MΩ |
To 6.0 MΩ |
< 2.7 V dc |
< 0.7 V dc |
< 0.9 V dc |
< 350 µA |
Diode Test |
< 2.7 V dc |
2.000 V dc |
< 1.2 mA |
General specifications |
Function |
Description |
Maximum voltage between any terminal and earth ground |
600 V |
Surge protection |
6 kV peak per IEC 61010-1 600V CAT III, Pollution Degree 2 |
W Fuse for A input |
11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293) |
Display |
Digital: 6,000 counts, updates 4/sec; Bar Graph: 33 segments, updates 32/sec |
Temperature |
Operating: -10°C to +50°C; Storage: -40°C to +60°C |
Humidity |
0% to 90% to 35°C 75% to 40°C 45% to 50°C |
Temperature coefficient |
0.1 x (specified accuracy/°C) (< 18°C or > 28°C) |
Operating altitude |
2,000 meters |
Battery |
9 Volt Alkaline, NEDA 1604A/IEC 6LR61 |
Battery life |
Alkaline: 400 hours typical, without backlight |
Safety compliances |
ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA C22.2 No 61010-1-04, UL 6101B (2003) and IEC/EN 61010-1 2nd Edition for measurement Category III, 600 V, Pollution Degree 2, EMC EN61326-1 |
Certifications |
UL, CSA, TUV, N10140 , VDE |
IP rating (dust & water protection) |
IP42 |